1. Provided with refrigration system to control temperature range from 10C to 60C +- 2C and humidity 55% to 95% +- 5%.
2. Temperature and Humidity can be controlled through Microprocessor Based Digital Temp. & Humidity Indicator cum controller.
3. Double wall, inner made of Stainless Steel,outer made of thick best quality Mild Steel with powder coated.
4. Insulation between inner and outer wall is done with glass wool to prevent any heat loss.
5. Inner chamber is provided with 3Nos. perforated Stainless Steel shelves with ribs to adjust the shelves at desirable height.
6. Provided with Air Circulation fan in the chamber for uniformity of temperature.
7. Timer 1minute to 99hrs.
8. Provided with double door,inner made of thick transparent full view glass and outer door is double walled metallic properly insulated with glass wool and also provided with best quality handle,gasket etc.
9. With water level indicator.
10. Whole unit is mounted on castor wheels.
11. Interior Chamber with flourescene tubes.
12. Shape & Size can be vertical in size with panel attached on top or side panel(as per client requirement).
13. Leakproof positive locking arrangement will be there.
14. CFC free ECO friendly compressor.
15. To work on 220/230volts A.C.

Description

 

Temperature Range : 20 DegreeC to 60 DegreeC Accuracy : +- 0.2 DegreeC
Humidity Range :40% RH to 95% RH Accuracy : +-2% RH
Temperature Uniformity : +-1 DegreeC Humidity Uniformity : +- 3% RH

 

  • Temperature & Humidity Suitable for 25 DegreeC& 60% RH, 40 DegreeC& 75% RH, 30 DegreeC& 65% RH
  • Specially designed chambers to carry out stability studies at low humidity conditions such as 25 DegreeC& 40% RH, 40 DegreeC& 25% RH NMT, 30 DegreeC& 35% RH

 

Chamber Size In Mms Inches Cap.
900x600x600 (36″x24″x24″) 336l

Humidity System

Humidity is by steam injection. Humidity system consist of boiler tank, which is fitted with two nos. of boiler heaters, one heater works as a stand by and automatic switch over to stand by heater if regular fails and give alarm.A water reservoir tank is connected to the boiler tank. DM water is preferred. A magnetic float switch provided for safety for boiler low level to cut off the its supply.

Alarms

Audio and visual alarms are generated for temperature / humidity variation, utility failure.

Safety

The unit is provided with safety devices for temperature, and humidity over shoot in case of malfunction.

Chamber Size 900x600x600 mm
Model/ Code HL-2111
Temperature Range 10 Degree C to 60 Degree C
Voltage 220/230 V
Country of Origin Made in India
Brand Physilab
Usage/Application For Laboratory